Used PLANMECA PM 2002 CC Proline Dental X-Ray SE de Vende - DOTmed Lista #1776308: 2 x Planmeca Machines Available, 1 x Gendex 1x Satelec Descriptions as follows: 1) Planmeca PM 2002 EC Proline Panaromic X-Ray Unit with Xfreedom Velopex X-Ray Photo Developer Serial Numbers: OPG: CJD403173 Photo Developer: XF50001R Associated Equipment: Film Autoprint. Service Technicians Dental Forum Planmeca Pm 2002 Cc OWALDO QUISPE PEREZ SESEL I need service manual panoramic rx planmeca PM2002 CC.
DESCRIPTION
The Proline PM 2002 is a dental panoramic film based x-ray.
FORUMSView All (1)
Ask a New Question1Replya year ago | a year agoService manual Can you help me with the service manual of the orthopantomograph planmeca 2002 please, I need it!!Reply |
FEATURES
- Pediatric, automatic double TMJ, vertical segmenting, and sinus exposure settings
- High frequency, constant potential DC generator
Additional Specifications
- Generator: DC, constant potential, micropreocessor controlled
- 80kHz
- 0.5 x 0.5mm focal spot
- 2.5 mm AI total filtration
- Anode Voltage: 60-80kV
- Anode Current: 4-12mA DC
- Exposure Time: 5-18s
- Film Size: 15 x 30cm; 12.5 x 30cm
- Cassette: flat
- SID: 480mm (19″)
- Magnification: constant 1.2
- Line Voltage: 100/117/220-230/240V; 50/60Hz
- Line Current: 8-16A
- Dimensions: 60″W x 40.5″D x 86.5″H
- Weight: 237lbs
Planmeca 2002 cc proline error
September 02, 2014 10:10
Hello, I've the Planmeca Proline 2002 cc Ortho and the machine reports me the Error 57 in the display. I've turn off and then turn on the Ortho, but the error continues.
Can anyone help me?
Thanks!
re: Planmeca 2002 cc proline error
September 03, 2014 09:01
remove the handswitch and turn unit off and on and see if it clears? could be a bad switch closed all the time
Log inor Registerto rate and post a comment
re: Planmeca 2002 cc proline error
September 03, 2014 09:01
Error 57 hand switch activated during switch on check this first
The Videojet 1520 small character inkjet coder is a highly flexible platform specifically designed for medium duty coding applications. Its advanced performance features make it a benchmark performer on packaging lines across the globe. The Videojet 1520. Download Videojet Operator Manuals Videojet offers a library of operator manuals for our printing technologies. Please verify your identity and choose the. Videojet 1520 manual download. Videojet 1620/1620 UHS Operator Manual Ink Core Life Videojet 1620/1620 UHS 14000 Hours Table 7-6: Ink Core Life Fault Messages Fault messages are displayed at the bottom left of the screen to supplement the printer fault icons. The messages are arranged so that, if more than one fault is found, only the most serious fault message is displayed. Videojet 1710 Operator Manual The ink is transferred from the ink cartridge to the ink core under two conditions:. Ink fill for new printer/new ink core commissioning. Ink fill after normal usage Note: Before filling ink, the user must make sure that the type of ink used is of pigmented variety and suitable for the nozzle fitted to the. Aug 22, 2017 Save this Book to Read videojet 1520 service manual PDF eBook at our Online Library. Get videojet 1520 service manual PDF file for free from our online library.
re: re: Planmeca 2002 cc proline error
September 03, 2014 12:54
This policy will: Communicate the Spa expectations for attendance standards. Attendance PolicySampleYOUR Spa Name time and attendance standards apply to all team members that are classified as non-exempt. Ispa spa operations manual download.
Is the hand switch the button you press to take a radiography? I don't have this piece instaled. I only have the autoprint machine instaled. Sometimes after the error 57 appears HE 2 it seems like the
Keyboard doesn't respond to the key press. Should I connect the handswitch? How can i install it?
Thank you!!
to rate and post a comment
re: Planmeca 2002 cc proline error
September 03, 2014 09:02
Hi, you have problem in exposure button, praticaly is switched on. Open the exposure botton and also you can have problem in the extention cable. Put out the cable of extention exposure button and the error must deasapeare.
Send me a mail, and i send you technical manual.
Boian
to rate and post a comment
re: re: Planmeca 2002 cc proline error
September 03, 2014 12:54
Is the hand switch the button you press to take a radiography? I don't have this piece instaled. I only have the autoprint machine instaled. Sometimes after the error 57 appears HE 2 it seems like the
Keyboard doesn't respond to the key press. Should I connect the handswitch? How can i install it?
Thank you!!
re: re: re: Planmeca 2002 cc proline error
September 03, 2014 05:34
HE 2 Amalgam container is 100% full (only Metasys
or Dürr).
Replace or empty the container immediately.
The suction system cannot be used before the
container is emptied or replaced. See the
User’s manual for details.
send me an email address I will send you the service manual, if your keyboard does not respond and is used to initate the exposure it probably is bad
to rate and post a comment
re: re: re: re: Planmeca 2002 cc proline error
September 04, 2014 10:19
Amalgam in x-ray?!?
Log inor Registerto rate and post a comment
Planmeca Pm 2002 Cc Service Manual
re: re: re: re: re: Planmeca 2002 cc proline error
Planmeca Promax Service Manual
September 05, 2014 09:38
that came from the service manual
re: re: re: re: Planmeca 2002 cc proline error
March 10, 2017 10:08
Hello Lenny, i need this service manual, can you send it also to me please [email protected]
Thanks so much
re: re: re: re: re: Planmeca 2002 cc proline error
March 10, 2017 10:33
OK I will send
Log inor Registerto rate and post a comment
re: re: re: re: re: re: Planmeca 2002 cc proline error
March 10, 2017 04:46
Planmeca Promax Installation Manual
Thanks a lot