Centrifuge 5427 R was designed with high-throughput research applications in mind. Choose the Centrifuge 5427 R if a 24-place microcentrifuge (e.g., Centrifuge 5424 R) doesn’t meet your needs regarding tube capacity and if you don’t need the extra versatility our cross-over Centrifuge 5430 R. Buy Eppendorf 022662201 SERVICE MANUAL 5804/R,5810/R at PartsSource. Largest Online marketplace for Medical Equipment Replacement Parts featuring OEM Biomedical parts. Medical parts online catalog – easy part ID and SmartPrice. Centrifuge 5804/5804 R/5810/5810 R — Operating manual 2.1 Main illustration Fig. 1: Depiction of Centrifuge 5810 and Centrifuge 5810 R. The Centrifuge 5804 and Centrifuge 5804 R are similar in design. 1 Centrifuge lid 2 Monitoring glass Visual control for rotor stop or option for speed check via stroboscope 3 Operator panel with display.

Advertising

Centrifuge 5804/5804 R/5810/5810 R — Operating manual

Eppendorf Centrifuge 5810 R User Manual

68

No rotor

Centrifuge does not start
up.

Error in the drive or in the rotor detection.

Switch the centrifuge off and back on
again after > 20 s.

Press Open

Centrifuge lid could not be locked.

1. Press the open key.
2. Try again to close centrifuge lid.

Close lid

Centrifuge lid not closed properly.

Close the centrifuge lid firmly.

Lift lid

The centrifuge lid does
not open.

The centrifuge lid cannot open automatically.

Lift the centrifuge lid manually.

IMBAL

The centrifuge shakes
when it starts up and
switches off.

Rotor asymmetrically loaded.

Load the rotor symmetrically (see p. 51).

ROTOR

The centrifuge shakes
when it starts up and
switches off.

Rotor not screwed sufficiently tight.

1. Tighten the rotor nut (see p. 51).
2. Check the rotor cone and motor shaft for

grooves and damage.

ROTOR

The centrifuge shakes
when it starts up and
switches off.

Eppendorf Centrifuge 5804 Manual

Centrifuge was pushed.

Table is not stable.

Position the centrifuge on a stable table
(see p. 48).

SPEED

Centrifuge switches off.

Nominal speed for rotor too high.

Enter the appropriate nominal speed
(see p. 13).

change rotor

The maximum service life of the rotor has
been reached. The warning is displayed after
98,000, 99,000 and 99,600 runs (3 times
after each run). After 100,000 runs, it is
displayed after every run.

Contact Technical Service.

Temperature display
flashes.
(only 5804 R/5810 R)

Temperature deviation from the set value: ±3
°C.

Check the settings.

Once installation is completed click 'OPEN'.5. Enjoy using 9Apps. If a prompt pops like below, click setting to continue to install.4. Sansui tv remote app download for android. It will show a warning on google chrome but it's completely safe to download.Click 'OK'.3.

Wait until the target temperature has been
reached.

Check unhindered air circulation through
the air slots.

Thaw ice or switch off device and allow it
to cool down.

Overtemp

(only 5804 R/5810 R)
Centrifuge switches off
and issues a warning
tone.

Temperature deviation from set value in the
rotor chamber: ±5 °C.

Check the settings.

Check unhindered air circulation through
the air slots.

Thaw ice or switch off device and allow it
to cool down.

Clear memory

Program memory full.

Eppendorf Centrifuge 5810 R User Manual

Delete some programs (see p. 61).

Interrupt

Power failure during a run.

Check the mains connection.

Error 1

Error in speed measuring system.

If this error message appears again, test
with a different rotor.

Error 2

Imbalance sensor faulty.

Repeat the run.

Symptom/message

Cause

Remedy

EN

Operating manual

Advertising
This manual is related to the following products:

5810 Centrifuge, 5804 R Centrifuge, 5804 Centrifuge

Popular Brands
Popular manuals
One moment please..
  • Documents
  • Parts
  • Videos
  • News
DESCRIPTION
Unparalleled versatility.

Featuring powerful, maintenance-free motors, these models accommodate a variety of rotors to meet your application needs. They are three centrifuges in one: a high-capacity, general-purpose centrifuge for cell harvesting; a high-speed centrifuge for separating cell lysates; and a microcentrifuge for DNA precipitations.
Model 5810 and refrigerated Model 5810 R* offer maximum application versatility and popular tube capacity, accommodating up to 4 x 400 ml tubes or 16 microplates. These compact, benchtop centrifuges achieve remarkable throughput of up to 48 x 15 ml or 20 x 50 ml conical (Falcon®) tubes

FORUMSView All (11)
Ask a New Question

3

Replies
-doowon
6 months ago
6 months agoError code 31
Hello, have some issue with error code 31. when it started it door latch is hunting and ends up with error code 31.I replaced two sensors on the door latch but did not help. changed PCB board with brand new same thing happens. Anyone have any ideas?Reply
-paul.blackaby_3666
a year ago
a year agoService Manual
Anyone able to supply a service manual for a 5810R or the method of changing the lid strut. Thankyou in advanceReply

2

Replies
-TechST
2 years ago
2 years agoProcedure
Hi to all, i need replace the main board of this device and from Eppendorf says this procedure can do it only the oficial service because when you replace this PCB is necesary adjust the imbalance sensor. If in this case the information it's true:How can i do that procedure? Is posible to explain me?Thanks in advance..Reply

FEATURES

  • Interchangeable swingbucket and high-speed fixedangle rotors
  • Spin up to 14,000 rpm, or 20,800 x g
  • Low-profile 11.2 in/28 cm access height for easy sample loading
  • Automatic rotor recognition senses rotor type to set maximum allowable speed—ensures safety
  • Fast, simple programming of time, temperature (refrigerated models), speed/ g-force and radius value
  • 34 user-defined programs
  • Adjustable-speed Short Spin button
  • Ten acceleration and braking rates
  • Simple operation
  • Speed, g-force and radius correction values can be entered
  • Values can be changed during centrifugation
  • Temperature range from –9 °C to 40°C, with Fast Temp function for rapid cooling of the centrifuge chamber
  • Standby cooling maintains temperature when not in use
  • Speed can be set from 200 rpm to maximum speed in increments of 10 rpm
  • Short Spin button
  • Selectable program memory for up to 34 individual user programs
  • Ten acceleration and deceleration rates for sensitive sample material

Eppendorf Centrifuge 5810r Service Manual

SERIES MODELS
Height13 in
Length 24 in
Weight150 lbs
Width21 in
Additional SpecificationsMaximum power requirement: 900 W

Eppendorf Centrifuge 5810 R Service Manual

Height13 in
Length 24 in
Weight218 lbs
Width28 in
Additional SpecificationsMaximum power requirement: 1,650 W